SAFETY FEATURES

1. When the operation mode is set to BLUE Mode and the power setting is set to Max the unit will
automatically reduce power if operated at these settings for more than 10 minutes. (The output display
reduces from Maximum to Minimum power.) If the safety feature is activated release foot control.
2. The control unit will beep to alert user if the power switch is turned on while the foot control is depressed.
3. When foot control is continuously depressed for 20 minutes or longer, the unit will beep in a continuous
tone and power down. Release the footswitch control. Turn the power switch off then back on again.

CLEANING AND STERILIZATION

Contents of this package are not provided sterile. The following cleaning and/or sterilization procedures are
recommended for your Hu-Friedy magnetostrictive scaler.
PROCEDURES FOR CLEANING CONTROL UNIT AND COMPONENTS
1. Turn off the power switch.
2. Remove the insert from the handpiece.
3. Carefully wipe all surfaces of the control unit, handpiece and handpiece cord, power cord, water tube,
foot control and cord which are contaminated by touch, contact with instruments or other inanimate
objects during dental treatment with a soft cloth using only approved hospital grade surface disinfectants
(EPA-registered, VAH / DGHM, RKI approval or CE marked).
4. Always follow the manufacturer's directions for disinfectant use.
5. The use of an abrasive wipe or cleaning solution will damage the surface finish of the control unit.
6. Phenols, iodophors, hypochlorite/household bleach and citric acid based disinfectants are not recommended.
NOTICE: The following websites provide additional information on safety and asepsis procedures:
http://www.osap.org/indec.cfm and http://www.cdc.gov/OralHealth/infectioncontrol/index.htm
CAUTION: Do not sterilize the control unit, power cord, foot control, water tube, irrigation connector,
handpiece, or handpiece cable.
PROCEDURES FOR HU-FRIEDY ULTRASONIC INSERTS (sold separately)
1. Clean, inspect and steam sterilize before each use.
2. For complete instructions refer to cleaning and sterilization instructions provided with the ultrasonic insert.