HOLEX Digital outside micrometer with disc measuring surface
de
6.4.3.
Set zero point in "Absolute" measuring mode
1. For measuring range 0-25 mm:
Place the measuring spindle (3) on the anvil (2) by turning the friction ratchet (9) clockwise until the friction ratchet (9) is
bg
released.
For measuring range from 25 mm:
Place the measuring spindle (3) on the setting dimension (included in the scope of delivery) by turning the friction
da
ratchet (9) clockwise until the friction ratchet (9) is released.
2. Button „ON/OFF - SET" Press and hold until "Set" flashes in the display.
6.4.4.
Set zero point in "Incremental" measuring mode
en
1. TButton „ABS/INC - UNIT" button briefly to switch to the "Incremental" measuring mode.
▶ Zero point for incremental measurement is set, "0.000" is displayed.
2. Button „ABS/INC - UNIT" button briefly to switch back to the "Absolute" measuring mode.
fi
▶ The zero point saved in the "Absolute" measuring mode is retained.
7.
Maintenance
fr
Intervall
Nach eigenen Vorschriften
it
7.1.
CHANGING THE BATTERY
ü Bügelmessschraube ist ausgeschaltet.
1. Batterieabdeckung abnehmen.
hr
2. Batterie CR2032 austauschen.
▶ Pluspol zeigt nach oben.
3. Batterieabdeckung einsetzen.
lt
8.
Troubleshooting
Fault
nl
Display zeigt „E 1"
Display zeigt „Exxxxx"
no
Display zeigt „E 2"
Display zeigt „E 3"
pl
Display zeigt „E 8"
pt
Messung ungenau
ro
Display gestört oder ausgefallen
sv
Keine Anzeige
Display flackert
wird angezeigt
sk
9.
Cleaning
-8. .
Remove dirt using a soft clean dry cloth. After contact with liquids, dry mechanical parts well. Do not use chemical, alco-
sl
holic, abrasive or solvent-based cleaning agents.
10. Storage
es
In Originalbox bei Temperaturen zwischen -20 °C und +60 °C lagern. Do not store close to corrosive or aggressive chemical
substances, solvents, heat, moisture or dirt.
11. Technical data
cs
Bügelmessschraube
hu
Spindle
22
Wartungsarbeit
Rekalibrierung
Cause
Messung überschreitet Anzeigebereich Nullpunkt neu setzen
Wert des Nullpunkts zu hoch
Bügelmessschraube gestört
Sensorfehler
Messflächen verschmutzt
Nullpunkt falsch gesetzt
Messspindel zu schnell bewegt
Starker Störeinfluss
Batteriespannung zu niedrig
ABS INC Set
.
8 . .
.
8 . .
.
8 88
Auszuführen von
Kundenservice Hoffmann Group
Remedy
Auf inkrementellen Modus wech-
Nullpunkt neu setzen
Batterie entnehmen und wieder ein-
Bügelmessschraube zum Kundendi-
Messflächen reinigen
Nullpunkt neu setzen
Messspindel langsamer bewegen
Batterie entnehmen und wieder einset-
zen
Batterie austauschen
in
Upright
seln
setzen.
enst senden