Important Safety Information
• While using this device, make sure that no mobile phone or any other electrical devices that emit
electromagnetic fields is within 12 inches (30cm). This may result in damage to the performance of
the device.
• Inspect the main unit and the parts before each use to ensure that parts are not damaged, the
device is assembled properly and it is operating normally.
• To prevent damage to the device, add the medication slowly. DO NOT allow the medication to
overflow the Medication Port.
• To prevent damage to the device, make sure that the Mesh Cap is correctly in place. If the Mesh
Cap is not properly closed, medication will leak out of the device.
• DO NOT operate the device at temperatures greater than +104°F (+40°C).
• Operate the device only as intended. DO NOT use the device for any other purpose.
• Changes or modifications not approved by Omron Healthcare will void the user warranty.
• Dry the parts by air drying only. DO NOT use any equipment to dry the device parts.
• To prevent damage to the device, DO NOT immerse the Main Unit in any liquid, DO NOT wash or
rinse any of the parts under strong running water, and DO NOT touch the mesh with your hand or
any object.
• To prevent damage to the device, DO NOT carry or leave the Medication Container filled with
medication or distilled water.
• DO NOT use household bleach to clean the device or parts; doing so will cause rust and damage
the Mesh Cap.
• To prevent damage to the device, DO NOT clean the Main Unit using abrasive cleaners or any type
of chemical.
• NEVER clean with benzene or thinner.
• Ensure that this device has acclimated to room temperature before nebulizing the medication.
Nebulizing the medication after an extreme temperature change could lead to less nebulization.
OMRON recommends waiting for approximately 2 hours for the device to warm up or cool
down when the device is used at an environment within the temperature specified as operating
conditions after it is stored either at the maximum or at the minimum storage temperature.
For additional information of operating and storage/transport temperature, refer to section 7.
6