Descargar Imprimir esta página

Xylem SI Analytics CT 72/P Manual De Instrucciones página 71

Termostatos transparentes

Publicidad

Idiomas disponibles

Idiomas disponibles

6.4.
Over and Sub temperature warning function
Over temperature
Sub temperature
Safety installations, warning functions
If the observance of a working temperature value >SETP< has to be supervised for
a sensitive temperature application, then set over and sub temperature warning
values. In the example below the SETPOINT 45 °C is surrounded by the values
OVERTMP 487 °C and SUBTEMP 43 °C. The electronics immediately register if
the actual temperature breaches one of the set limit values. The resulting reaction
is defined in a further menu item. (See chapter 6.3.
to shutdown function)
1. Press the key
2. By pressing the
>SUBTEMP<.
3. Press the
key. The integer digits flash
4. Change the values to 87. °C and/or 83. °C by pressing the
and confirm with the
The decimal digits flash and can be adjusted if desired.
Confirm once more by pressing the
See above examples.
The warning functions are only activated if the actual bath temperature
remains within the set limit values for 3 seconds after switch-on.
Recommendation:
Set the over temperature warning value >OVERTMP< 5 °C to 10 °C above the
working temperature setpoint.
Set the sub temperature warning value >SUBTMP< 5 °C to 10 °C below the
working temperature setpoint.
.
or
key select the menu >OVERTMP< or
key.
71
Switch-over from warning
and
key.
key

Publicidad

loading