10.4. Status messages
Status messages
00 MANUAL STOP
01 MANUAL START
02 REMOTE STOP
04 REMOTE START
10.5. Error messages
Error messages
-01 TEMP / LEVEL ALARM
-02 REFRIGERATOR ALARM
-03 EXCESS TEMPERATURE WARNING
-04 LOW TEMPERATURE WARNING
-05 TEMPERATURE MEASUREMENT
ALARM
-06 SENSOR DIFFERENCE ALARM
2
-07 I
C-BUS ERROR
-08 INVALID COMMAND
-09 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE
-10 VALUE TOO SMALL
-11 VALUE TOO LARGE
-12 WARNING : VALUE EXCEEDS
TEMPERATURE LIMITS
-14 EXCESS TEMPERATURE PROTECTOR
ALARM
-30 CONFIGURATION ERROR: CONFIRM
BY PRESSING <OK> ON CIRCULATOR
-33 SAFETY SENSOR ALARM
Remote control
Description
Thermostat in „OFF" state.
Thermostat in keypad control mode.
Thermostat in „r OFF" state.
Thermostat in remote control mode.
Description
Low liquid level alarm.
Control cable of the flow-through cooler is short-circuited or
interrupted.
High temperature warning.
Low temperature warning.
Working temperature sensor short-circuited or interrupted.
Sensor difference alarm.
Working temperature and safety sensors report a temperature
difference of more than 35 °C.
Internal error when reading or writing the I
Invalid command.
Invalid command in current operating mode.
Entered value too small.
Entered value too large.
Error in A/D converter.
Excess temperature protector alarm
The configuration of the circulator does not conform to its
present use.
Press
to automatically perform a single modification of
the configuration.
Excess temperature sensor short-circuited or interrupted.
96
2
C bus.