Over Temperature Alarm; Therapy Inactive Alarm; Mattress Not Connected Alert - arjo TheraKair Visio Manual Del Usuario

Sustitutivo de descanso
Ocultar thumbs Ver también para TheraKair Visio:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 203
Alarm
Unit is over-temperature
Alarm Condition
Pause
Alarm
Therapy in-active
Alarm Condition
Pause
Alert
Mattress not connected!
Connect mattress?
Yes
Alert
Mattress not connected!
Ready to start?

Over Temperature Alarm

This screen is displayed when the therapy unit
has overheated, this may be due to a blockage
or a high ambient temperature.
Check that the air vents on the system are clear
Power the system down and allow it to
Reset

Therapy Inactive Alarm

This screen is displayed when the system has
been left in an alert condition for more than
two minutes.
The alarm sounder may be paused for two
minutes by pressing the pause button.
Reset

Mattress Not Connected Alert

This screen is displayed when the system is
powered up before connecting the air supply
connector to the therapy unit.
Connect the air supply connector and press the
YES button to clear this alert.
Press NO to display the CPR Alarm screen.
No
This screen is displayed after pressing the YES
button on the previous screen.
Connect the air supply connector and press
YES to begin therapy and clear this alert.
Yes
28