Alarms And Alerts; System Integrity Alarm; Air Pressure Alarm - arjo TheraKair Visio Manual Del Usuario

Sustitutivo de descanso
Ocultar thumbs Ver también para TheraKair Visio:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 203

Alarms and Alerts

The alarm message screen will give details of the particular alarm trigger and suggest a course of
action.
When the cause of the alarm has been removed, press the reset button to resume therapy.
i
Alarm
Alarm Condition
Pause
Alarm
Lost system integrity
Please contact your nearest
service centre.
Alarm Condition
Alarm
Unable to achieve air pressure
Alarm Condition
Pause
C.P.R.
Reset
Pause
Reset
CPR Alarm Screen
This screen is displayed when the air supply
connector is disconnected.
When the connector is reconnected, press
the reset button to clear this alarm.
The alarm sounder may be paused for two
minutes by pressing the pause button.

System Integrity Alarm

This screen is displayed when the system
experiences a fatal error.
Contact your local Arjo subsidiary or agent for
more details.
The alarm sounder may be paused for two
minutes by pressing the pause button.

Air Pressure Alarm

This screen is displayed when the system air
pressure falls outside the set parameters.
This may be due to damage to the mattress
or to the therapy unit.
Check the mattress and hose for damage.
Contact your local Arjo subsidiary or agent for
more details.
The alarm sounder may be paused for two
minutes by pressing the pause button.
27

Publicidad

Tabla de contenido
loading

Tabla de contenido