SPECIFICATIONS
Dimensions (H x W x L) ............................................................................................................................................................................... 8.4"H x 3.5"W x 8.6"L (21.3 cm x 8.9 cm x 21.8 cm)
Weight ............................................................................................................................................................................................................................................ 4.95 ± 0.1 lb. (2.25 ± 0.05 kg)
Oxygen Output ................................................................................................................................................................................................................. 90% + 4/-3 over entire operating range
Oxygen Prescription Settings .................................................................................................................................................................................................................................................1 to 5
Operating Temperature Range .............................................................................................................................................................................................. +5 °C to +35 °C (+41 °F to +95 °F)
Operating Humidity Range ............................................................................................................................................................................................................ 15 % to 93 %, non-condensing
Operating Atmospheric Pressure Range ............................................................................................................ 700 hPa to 1060 hPa (approx. 10,500 ft. to 1300 ft below Sea Level @ 35°C)
Transport & Storage Temperature Range .......................................... - 25 °C (-4 °F) without relative humidity control to +60 °C (+140 °F) at a relative humidity up to 93%, non-condensing
Transport & Storage Humidity Range .............................................................................................................................................................................................. 15% to 93% non-condensing
Transportation and Storage Atmospheric Pressure (at nominal temperature and humidity) ........................................................................................................................ 640 hPA - 1060 hPA
Maximum Limited Pressure ................................................................................................................................................................................................................................................... 24 psi
Maximum Breathing Rate ................................................................................................................................................................................................................................................. 40 BPM
Sound Level .......................................................................................................................................................................................................................37.5 dBA typical at setting 2 (MDS-Hi)
Measured Sound Pressure Level (ISO 80601-2-69:2014) ...............................................................................................................................................................................................< 45 dBA
Measured Sound Power Level (ISO 80601-2-69:2014) ...................................................................................................................................................................................................< 53 dBA
Measured Sound Pressure Level Range of Alarm (IEC 60601-1-8:2012) .......................................................................................................................................................................> 59 dBA
Power Quality Requirements
AC Mains Input ........................................................................................................................................................................................................................... 100-240 VAC, 50/60 Hz, 140 VA
DC Input ................................................................................................................................................................................................................................................................ 13.8 to 17 VDC
The battery will charge as long as the DC input voltage is higher than the battery voltage. The device will charge the battery to full capacity if the DC input is at or above 13.8V DC.
OSD Set Points .............................................................................................................................................................................................................≥ 86% - Normal Oxygen Symbol (Green)
Device Classification ............................................................................................................................................................................................................ Class 2, Type BF Applied Part, IP22
The Drive DeVilbiss iGo2 Portable Oxygen Concentrator meets RTCA DO-160G Section 21 Category M and Section 20 Category T Airline Use.
Patented Breath Sensing with SmartDose Technology – US 8061353
Patented PulseDose Technology – US 4519387 and US 4457303
CAUTION
When moving the iGo2 POC from an extreme environment, allow time for the device to acclimate to the recommended operating environment. Operating your concentrator
outside the recommended operating environment may impact performance, cause damage, and will void the warranty.
Specifications subject to change without notice.
Audible Alerts:
• Low Battery
• Low Oxygen Output
• No Breath Detected
• High Temperature
• Unit Malfunction
TABLE 1 – STANDARD AND SMARTDOSE
Total Volume per
Setting
Minute (ml/min)
1
260
2
440
3
725
4
880
5
1014
5 SD
1100
TABLE 2 - OXYGEN FLOW EQUIVALENTS
PulseDose Bolus Volume (mL)
iGo2 Pulse Setting
1
2
3
4
5
NOTE– This comparison is theoretical, based on typical patient breathing pattern assumptions, and is only a recommendation. This is not intended to replace titration. The patient
should be titrated to the portable oxygen concentrator at rest and with activity.
Per ISO 80601-2-67
+/-15% at Standard Temperature and Pressure (tested @~933 hPa, 25 ˚C +/- 5 ˚C)
+/-25% over entire operating range
The above chart shows oxygen output delivered at each of the settings during normal use. When SmartDose technology senses the need for more oxygen, your oxygen setting will
temporarily be increased by up to 1 setting to help keep you active and saturated.
EN - 18
MODE PULSEDOSE
®
Standard and SmartDose Mode Oxygen Bolus Dispense Table
Bolus Volume
Bolus Volume
@ 15 BPM (ml)
@ 20 BPM (ml)
17.3
13.0
29.3
22.0
48.3
36.3
58.7
44.0
67.6
50.7
73.3
55.0
Equivalent Oxygen Flow Setting
@ 20 BPM
13.0
22.0
36.3
44.0
50.7
OUTPUT
®
Bolus Volume
Bolus Volume
@ 25 BPM (ml)
@ 30 BPM (ml)
10.4
17.6
29.0
35.2
40.6
44.0
(L/m)
iGo2 Oxygen Output
1.0
90% +4 / -3
2.0
90% +4 / -3
3.0
90% +4 / -3
4.0
90% +4 / -3
5.0
90% +4 / -3
< 86% - Low Oxygen Symbol (Yellow)
< 85% - Low Oxygen Symbol (Yellow) and Audible Alert
Bolus Volume
@ 35 BPM (ml)
8.7
7.4
14.7
12.6
24.2
20.7
29.3
25.1
33.8
29.0
36.7
31.4
Bolus Volume
@ 40 BPM (ml)
6.5
11.0
18.1
22.0
25.4
27.5
A-125B