Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.
3
How to use
Clean, disinfect and sterilize the product before using it for the
first time and before every subsequent use.
3.1
Ignition test
Before and during the ignition test do not touch any non-sterile
objects with the APC applicator / the APC probe.
You must not touch the ignition adapter or the beam during the
ignition test in any case.
Do not point igniting APC applicators / APC probes at the ignition
adapter for longer than 10 sec (continuous activation or several
consecutive activations), because it could otherwise be dam-
aged.
Only for HF surgical devices of the VIO Series: For this test use
only foot switches with shielded cables.
Only for HF surgical devices of the ICC Series: For this test use
only the neutral electrode settings "split neutral electrode" ( || )
or "any neutral electrode" ( | || ).
1. Switch on device(s).
2. Select the following device settings on the HF surgical de-
vice:
– On VIO devices: FORCED APC, 25 – 50 Watt
– On ICC 350, 300: SPRAY COAG, 40 – 60 Watt
– On ICC 200: APC-Mode (coded for by Display A. + two
digits for the power), 40 – 60 Watt
3. Connect the APC applicator / APC probe to the argon plasma
coagulation device.