1. Place the product in a suitable rinsing basket carefully. In do-
ing so make sure the product is not touching other instru-
ments or parts of instruments.
2. Start a tested program with the following properties:
– Thermal disinfection: 5 to 10 minutes at 90 to 93 °C
(with a tolerance in accordance with DIN EN ISO 17665),
A
≥ 3000.
0
– Final rinse with distilled or deionized water.
– Sufficient product drying.
3. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.
4.9
Check
1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
If damaged, do not use this product!
4.10 Packaging
1. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/polyethylene and/
or in a sterilization container.