Descargar Imprimir esta página

Abbott FreeStyle Libre3 Manual Del Usuario página 61

Publicidad

Idiomas disponibles

Idiomas disponibles

Immunity test: Proximity fields from RF wireless communications equipment; IEC
61000-4-3
IEC 60601 test level: See table below
Compliance level: Compliance to the tested levels
Electromagnetic environment – guidance: Portable RF communications equipment
(including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the System, including cables
specified by Abbott Diabetes Care. Otherwise, degradation of the performance of
the System could result.
The table below lists the immunity test levels at specific test frequencies for testing
the e"ects of some wireless communications equipment. The frequencies and
services listed in the table are representative examples in healthcare and in various
locations where the System may be used.
a)
For some services, only the uplink frequencies are included.
b)
The carrier is modulated using a 50% duty cycle square wave signal.
c)
As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because, while it does not
ART42490-004_rev-A 10:30am 04-May-22

Publicidad

loading