1) Patient Key: main identification keys for
patient demographics, used for patient
demographic management, default:
PatientId.
2) Patient Info Source: specifying the data
source for patient demographic query,
default: DB.
3) Patient Demographic Notification to
EMR: switch for Gateway outbounding
patient demographic notification to EMR,
default: unchecked.
 Realtime Data Related: Gateway outbounds
device data in real-time to EMR periodically.
The central station outbounds periodical data
as per its own period.
1) Outbound Periodic Data: switch for
Gateway outbounding periodic data to
EMR, default: checked.
2) Periodic Data Contains NIBP: switch for
periodic data containing NIBP, default:
checked.
3) Periodic Data Outbound Interval: interval
time for periodic data outbounding, deault:
60s.
Vista 120 Gateway Instructions for use
4) Outbound NIBP: switch for outbounding
NIBP unsolicitedly, default: checked.
5) Outbound QuickTemp: switch for
outbounding QuickTemp unsolicitedly,
default: checked.
 Alarm Switch Config: the configuration items
for Gateway outbounding alarms to EMR.
1) Parameter Alarm: switch for outbounding
parameter alarm message, default:
unchecked.
2) Arrhythmia Alarm: switch for outbounding
arrhythmia alarm message, default:
unchecked.
3) Life Alarm: switch for outbounding Life-
threatening alarm message, default:
unchecked.
Configuration