Valor 5000 Series
Span Calibration Procedure
• When CAL is displayed, press Yes to enter the calibration
sub-menu.
• When Span is displayed, press Yes to begin the Span
calibration process.
• Clear the pan if prompted to do so, then press Yes. . . . .
• The display will show "--C--", followed by the span
calibration weight value.
• If desired, press No to change to an alternate span
calibration weight value.
• Place the indicated calibration weight on the scale and
press Yes.
• The display will show "--C--", then "done" and then
revert to weighing. Remove the calibration weight.
Note: The calibration procedure can be canceled at any
time by pressing the Exit button or turning the scale off.
Linearity Calibration Procedure
• When LIN is displayed, press Yes to begin the linearity
calibration process.
• Clear the pan if prompted to do so, then press Yes.
• The display will show "--C--", followed by the first
calibration weight value.
• If desired, press No to change to an alternate first
calibration weight value.
• Place the indicated calibration weight on the scale
and press Yes.
• The display will show "--C--", followed by the second
calibration weight value.
• Place the indicated calibration weight on the scale
and press Yes.
• The display will show "--C--", then "done" and then
revert to weighing. Remove the calibration weight.
Note: The calibration procedure can be canceled at any
time by pressing the Exit button or turning the scale off.
End
• Press Yes to leave this sub-menu and go to the next.
• Press No to return to the top menu item in this sub-menu.
EN-14