Valor 5000 Series
Error Code
Err 1 (Overload)
Err 2 (Underload)
Err 9 (Checksum error)
Err14 (Zero error)
Err 21 (Checksum error)
CAL E (Calibration error)
4.4 Service Information
If the Troubleshooting section does not resolve or describe your problem, contact your authorized
Ohaus Service Agent. For Service assistance in the United States and Canada, please call Ohaus
Corporation toll free at (800) 526-0659 from 8 AM to 5 PM Eastern Time. An Ohaus Product Service
Specialist will be available to help you.
4.5 Accessories
Description
Foot Switch
TABLE 4-2. ERROR CODES.
Cause
Load exceeds capacity.
Missing pan.
Configuration data
checksum error.
Zero limit exceeded
during power up.
Calibration data
checksum error.
Wrong calibration weight
values used.
Remedy
Reduce load.
Install the pan and press the zero
button.
Service required.
Remove load from pan.
Check menu settings and recalibrate.
If error persists, service is required.
Calibrate using correct calibration
weights.
EN-22
Part Number
71169633