Applicable Products:
DEscription oF proDuct: A pelvic holder to prevent sliding. For chair use only.
4125c
posEY soFt BElt
4125C Chair use only; 4½"W x
REF
16½" (11 cm x 42 cm) belt
pad w/6 foot (2 m) straps
ApplicAtion instructions:
Make sure patient wears proper undergarments
to protect skin.
Before use, check device for damage. Discard if you
have any questions about patient safety.
1. Lay the device on the chair with the narrow side to the back
of the chair.
2. Bring the ends of the connecting straps on
the narrow end, down behind the chair, and
secure out of the patient's reach (fig. 1).
3. Position the patient as far back in the
seat as possible, with the buttocks
against the back of the chair.
4. Bring the wide part of the pelvic holder up
between the patient's legs.
5. Bring the ends of the connecting straps
on the wide end, down between the seat
and wheelchair sides at a 45-degree angle
(fig. 2).
6. Criss-cross the straps and use quick-
release ties to attach the straps to the
opposite side kick spurs, out of the
patient's reach (fig. 3).
7. If the chair has an adjustable seat, secure
straps to a movable part of the chair
frame, out of the patient's reach.
8. Check that the straps are secure and will not change position,
loosen, or tighten if the patient pulls on them, or if the chair
is tilted or adjusted.
ApplicAtion instructions: (plAstic incontinEncE
shiElD):
1. Insert the pelvic straps through the shield starting at the
widest part and pull the shield up until it rests on the foam
pelvic piece.
© 2009 Posey Company. All rights reserved.
4125c
REF
inDicAtions For usE:
• Patients assessed to be at risk of injury from a fall.
• Patients needing a positioning device for added safety while in a chair.
• Patients who have a tendency to slide down in a chair.
contrAinDicAtions:
• DO NOT use on a patient who is or becomes highly aggressive, combative, agitated, or
suicidal.
• DO NOT use on patients with: ostomy, colostomy, or G-tubes; hernias, severe Cardio
Obstructive Pulmonary Disease (COPD); or with post-surgery tubes, incisions, catheters, or
monitoring lines. These could be disrupted by a restraint.
ADVErsE rEActions
• Severe emotional, psychological, or physical problems may occur: if the applied device is
uncomfortable; or if it severely limits movement. If symptoms of these problems ever appear
for any reason, get help from a qualified medical authority and find a less restrictive, product or
intervention.
FiG. 1
FiG. 3
FiG. 4
Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com
Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014
Heed these warnings to reduce the risk of serious injury or
death:
• Monitor skin conditions in the groin area frequently. If the
patient slides down or forward, pelvic straps may damage
the skin.
• There is a risk of chest compression or suffocation, if the
patient's body weight is suspended off the chair seat. Use
extreme caution with chair cushions. If a cushion dislodges,
straps may loosen and allow the patient to slide off the
seat (fig. 4).
• Monitor per facility policy to ensure that the patient
cannot slide down, or fall off the chair seat and become
suspended (fig. 4).
45 ˚
FiG. 2
• STOP USE AT ONCE: if the patient has a tendency to slide
forward or down in the device; or is able to self-release.
ApplicAtion
instructions
rx onlY
FiG. 4
ADDitionAl sAFEtY AnD lAunDErinG
instructions on othEr siDE