Descargar Imprimir esta página

ResMed AirSense11 Guia Del Usuario página 31

Publicidad

Idiomas disponibles

Idiomas disponibles

Sound power level measured according to ISO 80601-2-70:2015 (CPAP mode):
Device with SlimLine and HumidAir 11 tub (HumidAir 11
tub 1/2 filled)
Device with SlimLine and Side cover
Physical Dimensions
Dimensions (H x W x D) with HumidAir 11 tub:
Dimensions (H x W x D) with side cover
Air outlet:
Weight - device and HumidAir 11 tub:
Weight - device with side cover
Housing construction:
Hot plate - Material:
Water capacity:
Time between each refill of the humidifier tub:
Recommended water type to use in the humidifier tub
(Standard tub):
Recommended water type to use in the humidifier tub
(Cleanable tub)
Humidifier tub - Material:
65W power supply unit
AC input range
DC output
Typical power consumption
Peak power consumption
Class of equipment
Environmental conditions
Operating temperature
Operating humidity
Operating altitude
Storage pressure/Storage altitude
Storage and transport temperature
Storage and transport humidity
Air Filter
Standard:
Hypoallergenic:
22
35 dBA with uncertainty of 2 dBA
33 dBA with uncertainty of 2 dBA
3.72" x 10.21" x 5.45"
(94.5 mm x 259.4 mm x 138.5 mm)
3.72" x 9.32" x 5.45"
(94.5 mm x 236.8 mm x 138.5 mm)
The 22 mm conical outlet connector complies with EN ISO 5356-1:2015
40 oz (1130 g)
41 oz (1142 g)
Flame retardant engineering thermoplastic
Stainless steel
380 mL
> 8 hours ±0.5 hours (tested at 23 ±2°C / 73.4 ± 3.6 °F)
Distilled water
Drinking quality water (potable)
Injection molded plastic, stainless steel and silicone seal
100-240V, 50-60Hz, 2.0A
115V, 400Hz, 1.5A (for aircraft use)
24 V
2.71A
56.1W (111.5VA)
73.2W (137.6VA)
Class ll
+41°F to +95°F (+5°C to +35°C)
Note: The airflow for breathing produced by this therapy device can be
higher than the room temperature. Under extreme ambient temperature
conditions (104°F/40°C) the device remains safe.
10 to 95% relative humidity, non-condensing
Sea level to 9,870' (3,010 m); air pressure range 1060 hPa to 700 hPa
1060 to 700 hPa relative humidity, non-condensing
-13°F to +158°F (-25°C to +70°C)
5 to 95% relative humidity, non-condensing
Material: Polyester non woven fiber
Average arrestance: >75%, when tested to EN779.
Material: Blended synthetic fibers in a polypropylene carrier
Efficiency: >80% (average) when tested to EN13274-7.
Note: The use of a ResMed approved hypoallergenic filter will result in a
small reduction in the accuracy of the delivered pressure at high leaks.

Publicidad