Technical Data
GB
Instrument
Light source:
Wavelength accuracy
Photometric accuracy*
Photometric resolution
Power supply
Operating time
Auto-OFF
Display
Storage
Interface
Time
Calibration
Dimensions
Weight
Ambient conditions
Waterproof
*measured with standard solutions
To ensure maximum accuracy of test results, always use the reagent systems supplied by
the instrument manufacturer.
38
Technical Data
triple
wavelength, automatic wavelength selection,
direct reading colorimeter
LEDs,
interference filters (IF) and photosensor in transparent
cell chamber. Wavelength specifications of the IF:
430 nm
= 5 nm
530 nm
= 5 nm
560 nm
= 5 nm
± 1 nm
3% FS (T = 20° C – 25° C)
0.01 A
4 batteries (AA/LR6)
53hr operating time or 15000 test measurements in continuous
mode when display backlight is off
automatic switch off
10 minutes after last keypress
backlit LCD (on keypress)
internal ring memory for 16 data sets
IR interface for data transfer
real time clock und date
user and factory calibration
resetting to factory calibration possible
190 x 110 x 55 mm (LxWxH)
approx. 455 g (incl. batteries)
temperature: 5 – 40 °C
rel. humidity: 30 – 90 % (non-condensing)
as defined in IP 68 (1 hour at 0.1 meter); floatable instrument