Progress through list is controlled by the trigger system. If trigger source is internal or immediate, the dwell time
setting (LIST:DWELl) determines time spent at each frequency. For any other trigger source, dwell time is
determined by trigger event spacing.
Burst Mode
The instrument can output a waveform for a specified number of cycles, called a burst. Burst is allowed with sine,
square, triangle, ramp, pulse, PRBS, or arbitrary waveforms (noise is allowed only in gated burst mode; DC is not
allowed).
For details, see Burst.
To Select Burst
Burst cannot be enabled when sweep or modulation is enabled. Enabling burst turns off sweep and modulation.
To avoid multiple waveform changes, enable burst mode after configuring other parameters.
Front Panel Operations
Press [Burst] > Burst ON |
SCPI Command
[SOURce[1|2]:]BURSt:STATe {ON|1|OFF|0}
Burst Mode
Burst has two modes, described below. Selected mode controls allowable trigger source, and which other burst
parameters apply.
– Triggered Burst Mode (default): The instrument outputs a waveform for specified number of cycle (burst count)
each time trigger is received. After outputting specified number of cycles, instrument stops and waits for next
trigger. The instrument can use an internal trigger to initiate burst, or you can provide external trigger by pressing
the front panel [Trigger] key, applying trigger signal to front panel Ext Trig connector, or sending software trigger
command from remote interface.
– External Gated Burst Mode : Output waveform is on or off, based on level of external signal applied to front panel
Ext Trig connector. When the gate signal is true, the instrument outputs a continuous waveform. When the gate
signal goes false, the current waveform cycle is completed and the instrument stops while remaining at the voltage
level corresponding to the starting burst phase of the selected waveform. The noise waveform output stops
immediately when the gate signal goes false.
Parameter
Burst Mode
(BURS:MODE)
Triggered Burst Mode:
TRIGgered
Internal Trigger
Triggered Burst Mode:
TRIGgered
External Trigger
Keysight EDU33210 Series User's Guide
OFF
> Burst
ON
| OFF .
Burst Count
(BURS:NCYC)
Available
Available
Burst Period
Burst Phase
(BURS:INT:PER)
(BURS:PHAS)
Available
Available
Not Used
Available
TriggerSource
(TRIG:SOUR)
IMMediate
EXTernal, BUS