Descargar Imprimir esta página

Keysight Technologies EDU33210 Serie Guia Del Usuario página 152

Publicidad

Idiomas disponibles

Idiomas disponibles

An LFSR is specified by the number of stages it contains and which stages ("taps") feed the exclusive-or (XOR) gates
in its feedback network. The PRBS output is taken from the last stage. With properly chosen taps, an L-stage LFSR
produces a repetitive PRBS of length 2
PRBS.
You can set L to 7, 9, 11, 15, 20, or 23, resulting in sequences from 127 to 8,388,607 bits in length.
The default value for L is 7, resulting in a sequence of 127 bits in length.
Modulation
Amplitude Modulation (AM)
The instrument implements two forms of AM:
– Double-sideband full-carrier (DSB-FC), which has an ITU designation of A3E and is used in AM broadcasting.
The equation for DSB-FC is
y(t)= [(½)+(½)•d•m(t)]•A
where
m(t) is the modulating signal
A
is the carrier amplitude
c
ω
is the carrier frequency of the carrier
c
d is the "modulation depth," or fraction of the amplitude range is used by the modulation
For example, a depth setting of 80% varies the amplitude from 10% to 90% of the amplitude setting (90% - 10% =
80%) with either an internal or a full-scale (±5 V) external modulating signal. You may set depth as high as 120%, as
long as you do not exceed the instrument's maximum output voltage of (±5 V into 50 Ω, ±10 V into high
impedance).
The top trace below represents the modulating signal; the bottom trace represents the modulated carrier.
152
L
- 1. The clocking frequency of the LFSR determines the "bit rate" of the
•sin(ω
t)
c
c
Keysight EDU33210 Series User's Guide

Publicidad

loading