3000 Series Indicators
3.4
Setup Menu
When the Indicator is used for the first time, enter this
menu to set the Capacity and Graduation.
3.4.1 Reset
Reset the Setup menu to the factory defaults.
No
= not reset.
Yes
= reset.
NOTE: If the Legal for Trade menu item is set to ON, the Capacity, Graduation, Zero Range and Legal For
Trade settings are not reset.
3.4.2 Legal for Trade
Set the legal for trade status.
OFF
= off
ON
= on
Turning on the "LFT" menu setting has the following effects:
• Zero-range is set and locked on "2".
• Auto Zero Tracking is set and locked on 0.5d
• The lb:oz unit is not available as a power-on setting.
3.4.3 Calibration Unit
Set the unit during calibration.
CAL UN kg = Calibrate using kg weights
CAL UN lb = Calibrate using pound weights
3.4.4 Capacity
Set the scale capacity from 5 to 20000. Refer to the Setup Table 3.3 for available settings.
Reset
No, Yes
Legal for Trade
Off, On
Cal Unit
kg, lb
Capacity
5...20000
Graduation
0.001...20
Power On Unit
g, kg, lb, oz, lb:oz, Auto
Zero Range
0%, 2%, 100%
End Setup
Exit SETUP menu
EN-19