SPECIFICATIONS
ELECTRICAL
Input Voltage: 100 - 240 VAC, 50 - 60 Hz
Input Current: 1A (@ 100 VAC)
Output: 60 VDC @ 0.4A
Leakage Current: less than 100 microamps
DIMENSIONS
11.43 cm (W) x 16.5 cm (L) x 6.35 cm (H)
Weight Under 5 Lbs.
ILLUMINATION SYSTEM S6LARC'"
Welch Allyn Lamp 09500
Color Temperature: 5500°K
DIMMING SYSTEM
Manual Shutter: 20% to 100%
FIBER OPTIC BUNDLE INTERFACE
Standard ACMI Connector
A P P R O V A L S
C E
IEC 601-1
UL 2601-1
The CE mark on this product indicates it has been tested to and conforms
with the provisions noted within the 93/42/EEC
ENVIRONMENT
Storage:
Operating: Humidity: 10% to 95% non-condensing
EQUIPMENT CLASSIFICATION
Class I, Type B, Continuous Operation, Ordinary Equipment, not
suitable for use in the presence of flammable anesthetic mixtures
FUSES
T1.0A L250V: 1.0A, 250V,
low breaking capacity
PATENTS
This product is covered by the following patents:
5083059; 5117154; 5138228; 5144201; 5291100; 5594826
CAN/CSA C22.2 No. 601.1
EN 60601-1-2
EN 60601-1
Humidity: 10% to 100%
Temperature: -25" C to +60" C
Pressure: 500hPA to 1060hPA
Temperature: +15" C to +30" C
Lamp: 60 Volts, 24 watts
Lamp Life: 350 hours average
Medical Device Directive.