12.6 Gas settings
Gas settings are not used in Polytron 8900 UGLD.
13
Factory default settings
13.1 Fixed settings for relays, LED and
alarm
Setting
Fault relay
Energized / Normally Open
Yellow Fault LED
Lit if a fault message is available
Red Alarm LED
Single blink if A1 condition is present.
Double blink if A2 condition is
present. If an alarm is configured as
acknowledgeable and the alarm is
acknowledged, single/double blink
changes into continuously lit.
Alarm Hierarchy
A2 overrides A1 on the LED. However,
the A1 and A2 relays operate inde-
pendently. I.e. if A1 is acknowledgeable
and A2 is not, and the gas concentra-
tion is such that it triggers A1 and A2:
acknowledging will cause the A1 relay
to release. However, the red LED will
still double blink as long as the A2 con-
dition continues to exist.
13.2 Settings which can be changed via
the menu
Menu
Default setting
Relay active on
Active on alarm
alarm / No alarm
A1 alarm: direction Rising
A2 alarm: direction Rising
A1 latch mode
Non-latching
A2 latch mode
Latching
A1 acknowledg-
Acknowledgeable
ment mode
A2 acknowledg-
Not acknowledge-
ment mode
able
Normal operation
Energized
A1 relay
Normal operation
A2 relay
UGLD signals
Alarm normal
Calibration pass-
_ _ _ 1
word
Settings password _ _ _ 2
Instructions for use
|
Dräger Polytron
Range
On / Off
Falling / Rising
Latching / Non-
latching
Acknowledgeable /
Not acknowledge-
able / Pre-acknowl-
edgeable
Energized / Not
energized
Alarm normal /
Alarm enhanced /
Alarm only
®
8900 UGLD Firmware version ≥ 4.0.x
Menu
Default setting
LCD setting
On
SIL status
Off
Language
EN
Function key
Faults
Interface
Default setting
Fault current
1.2 mA
Warning
Off
Warning current
3.0 mA
Warning cycle time
9 s
T1 (measurement
current)
Warning cycle time
1 s
T2 (warning cur-
rent)
Maintenance sig-
static
nal
Maintenance cur-
3.4 mA
rent
Analog offset
0 mA
Analog span
0 mA
HART address
0
13.3 Sensor specific values
Menu
Default setting
A1 Alarm
108 dB
A2 Alarm
108 dB
Hysteresis for A1
0
alarm at direction
rising
Hysteresis for A2
0
alarm at direction
rising
Delay
30 s
UGLD delay: Dis-
disabled
play
UGLD delay:
disabled
Alarm Relays
UGLD delay: 4-20
disabled
mA
Calibration inter-
3650
val [days]
Measuring range
< 54 dB
to
108.6 dB
|
Factory default settings
Range
On / Off
On / Off
DE / EN / FR / ES /
RU / ZH
Graph, fault, warn-
ing, vitality, bump
test
Range
0 to 3.5 mA
On / Off
0 to 3.5 mA
5 to 60 s
1 to (T1-1) s
static / dynamic
0 to 3.5 mA
-0.5 to 0.5 mA
-0.5 to 0.5 mA
0 to 15
Range
55 to 108 dB
0 to A1
0 to A2
1 to 60 (240) s
0 to 5475
not adjustable
en
23