Aesculap
®
Double pedal foot switch
4.1.2 Connect the double pedal foot switch
Risk of burns to the patient
caused by inadvertent activa-
tion of the Aesculap device
WARNING
when connecting a defective
double pedal foot switch!
There are two arrows given as markings, one on the HF
foot switch plug and the other on the HF foot switch
connection socket of the Aesculap device. The arrows
must point to each other when connecting the single
foot switch, see Fig. 1.
6
Always carry out a func-
tion check prior to each
use of the product.
Connect the HF cable and
the instrument only after
connecting the double
pedal foot switch.
Plug in the foot switch plug at the foot switch con-
nection socket of the Aesculap device, see Fig. 1.
Fig. 1
HF foot switch plug in connection socket HF
foot switch