It is advisable to use a different cup for each type of material to be mixed
(plaster, silicone, coating) as small amounts of residue from previous mixings can
compromise the final result.
It is also advisable to pre-mix the material manually with a spatula.
It is forbidden to mix materials different from dental model coatings or plasters.
It is forbidden to mix any materials in quantities that exceed the rated container
capacity. It is forbidden to use containers or spatulas that have not been approved by
the Manufacturer.
9
11
Figure
12
Figure
pressure
15
push stop
13
Figure
If, after the 3 end-of-programme beeps, the user fails to press the STOP key another 3
beeps are emitted 15 seconds later. If, once again, the user does not press the STOP
key, after another 30 seconds a continuous warning beep is emitted.
122 of 264
Fig. 11: Close the container: make sure that both the
edge of the cup (10) and the edge of the lid (9) are
thoroughly clean.
10
Fig. 12: Attach the cup to the machine: a magnetic
sensor will detect the presence of the cup/lid and start
the vacuum pump. After just a few seconds the cup will
be held in place automatically by the vacuum and the
display will show the achieved level of vacuum. Once the
vacuum is sufficient mixing begins: during this phase the
remaining mixing time is displayed as a 'countdown' .
Small variations in the strength of the vacuum do not in
any way compromise the final result.
Fig. 13: At the end of the cycle the machine emits 3
beeps: the words "press STOP key" flash on the screen.
Press the STOP (4) key.
The vacuum will be released and the cup will detach
4
from the machine.
MAINTENANCE AND USE MANUAL
EN
It is necessary to support the cup by
hand at the moment in which the mixing
cycle is interrupted.
ELite mix