8. Display messages
MEGABRAS DT70KV
When turning the equipment On using the On/Off
switch, this
introductory message appears for a while. During that time, the
equipment carries out some functional checking.
AUTO CHECK
Indicates that the equipment is performing internal checks.
PRESS START
Indicates that the equipment is able to start a test, for which the operator
must press the key
.
GENERATION LOCKED
If the control unit is turned on without being properly connected to the
generating unit, the message GENERATION LOCKED will be displayed
intermittently on the last line of the display.
If the Generation Lock button on the high voltage generating unit is
triggered this message is also displayed. In this case, unplug the machine
from the mains, release the Generation Lock button and reconnect the
dielectric tester to the mains.
INTERNAL PROTECTION ACTIVATED
This message indicates a malfunction on some of dielectric tester's
generation component. In this case the equipment should be sent to the
technical service.