Guidance and manufacturer's declaration – electromagnetic immunity
The Empi 300PV is intended for use in the electromagnetic environment specified below. The customer or the
user of the Empi 300PV should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
NOTE U
is the a.c mains voltage prior to application of the test level.
T
IEC 60601
test level
±6kV contact
±8kV air
±2kV for power supply
lines
±1kV for input/output
lines
±1kV differential mode
±2kV common mode
<5% U
T
(>95% dip in U
) for 0.5
T
cycle
40% U
T
(60% dip in U
) for 5
T
cycles
70% U
T
(30% dip in U
) for 25
T
cycles
<5% U
T
(>95% dip in U
) for 5 sec
T
3 A/m
Compliance level
±6kV contact
±8kV air
Not Applicable-
Battery Operated Device
Not Applicable-
Battery Operated Device
Not Applicable-
Battery Operated Device
3 A/m
40
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.