6. Model Orientation in the Stratos 300
6.5 Personalized model transfer with
the registration joint holder
6.5.1
Return articulator to the initial set position.
– See 6. 1 . 1 for description
Remove incisal indicator (1.2).
6.5.2 Mount the plane
indicator (50) instead of
the incisal guide rod (1.2).
6.5.3 Mount bite fork
support (51) to the lower
frame (2.2) by means of the
instrument carrier (40).
6.5.4 Insert registration
joint holder (52) instead of
the incisal plate (1.1) and
secure it.
6.5.5 Mount UTS
registration to the
registration joint holder
(52) and secure bite fork
with the bite fork support.
Example: Centric Tray
Example: UTS bite fork
(dentulous case)
6.5.6 Mount maxillary
model to the bite fork and
cast. Close articulator until
the plane indicator (50) lies
on the registration joint
holder (52).
Example: UTS bite fork
(edentulous case)
16
50
51
40
52
FH registration joint holder (54)
The procedure is the same
as with the CP registration
joint holder. The plane
indicator (50) lies on the
relevant support.
6.5.7 Subsequently,
orientate mandibular model in the usual manner.
50
54