9.5 Calibration
Benning guarantees compliance with the technical and accuracy specifications
stated in the operating manual for the first 12 months after the delivery date.
To maintain the specified accuracy of the measurement results, the instrument
must be recalibrated at regular intervals by our factory service. We recommend
a recalibration interval of one year. Send the appliance to the following address:
Benning Elektrotechnik & Elektronik GmbH & CO. KG
Service Centre
Robert-Bosch-Str. 20
D - 46397 Bocholt
9.6 Spare parts
Fuse F 15 A, 600 V, 50 kA, D = 10.3 mm, L = 38.1 mm, P.no. 10149447
10. How to use the protective rubber holster
-
The safety measuring leads can be stored by coiling them round the protec-
tive rubber holster M and clipping the probe into the holster M so that they
are sufficiently protected (see fig. 15)
-
You can clip one lead onto the protective rubber holster M in such a way that
the measuring probe projects. This allows you to bring the measuring probe
and the BENNING MM 5-1/ MM 5-2 up to the measuring point together.
-
The support at the back of the holster M can be used to prop the
BENNING MM 5-1/ MM 5-2 up in a diagonal position (to make reading
easier) (see fig. 16).
-
The protective rubber holster M is provided with a magnet which can be
used for suspension.
See fig. 15:
See fig. 16:
11. Technical data of the measuring accessories
-
Standard: EN 61010-031,
-
Maximum rated voltage to earth () and measuring category:
With push-on caps: 1000 V CAT III, 600 V CAT IV,
Without push-on caps: 1000 V CAT II,
-
Maximum rated current: 10 A,
-
Protective class II ( ), continuous double or reinforced insulation,
-
Contamination class: 2,
-
Length: 1.4 m, AWG 18,
-
Environmental conditions:
Maximum barometric elevation for making measurements: 2000 m,
Temperatures: 0 °C to + 50 °C, humidity 50 % to 80 %
-
Only use the measuring leads if in perfect and clean condition as well as
according to this manual, since the protection provided could otherwise be
impaired.
-
Throw the measuring lead out if the insulation is damaged or if there is a
break in the lead/ plug.
-
Do not touch the bare contact tips of the measuring lead. Only grab the area
appropriate for hands!
-
Insert the angled terminals in the testing or measuring device.
12. Environmental notice
At the end of the product's useful life, please dispose of it at
appropriate collection points provided in your country.