Never open the device. The device must not be modified, dismant-
led or repaired by the user. Repairs may only be carried out by
authorised specialists.
Never inflate the cuff when it is not properly applied to the upper
arm.
Use the device only with the approved upper arm cuff. The device
may otherwise sustain external or internal damage.
The cuff tube may only be removed from the unit by pulling on the
corresponding connector plug. Never pull on the tube itself!
Do not expose the device to extreme temperatures, humidity, dust
or direct sunlight because this may cause it to malfunction.
Keep the packaging, batteries and device out of the reach of
children.
Please comply with the storage and operating conditions defined
in Chapter 11 'Technical data'. Storage or use outside the specified
temperature and humidity range can affect the accuracy of the
measurement or the function of the device.
Do not use the device near strong electromagnetic fields and keep
away from radio equipment or mobile phones. Portable and mobile
high-frequency and communication devices, such as telephones
and mobile phones, can impair the functionality of this electronic
medical device.
Notes on electromagnetic compatibility
The device is suitable for use in all environments listed in these
instructions for use, including domestic environments.
The use of the device may be limited in the presence of electro-
magnetic disturbances. This could result in issues such as error
messages or the failure of the display/device.
Avoid using this device directly next to other devices or stacked
on top of other devices, as this could lead to faulty operation. If,
however, it is necessary to use the device in the manner stated,
this device as well as the other devices must be monitored to
ensure they are working properly.
The use of accessories other than those specified or provided by
the manufacturer of this device can lead to an increase in electro-
magnetic emissions or a decrease in the device's electromagnetic
immunity; this can result in faulty operation.
Failure to comply with the above can impair the performance of
the device.
Instructions for the measurement function check
Every Veroval
®
device has been carefully tested by HARTMANN for
measurement accuracy and has been developed with a view to a long
service life. We recommend carrying out a metrological check every 2
years for devices in professional use, for example, in pharmacies,
medical practices or hospitals. Please also observe the national regu-