Installation of the Unit
Factory Settings
The factory settings of the unit are 20 seconds
operation time.
Initial Steps
Charger
Please ensure first that the voltage stated on the
rating plate corresponds to the existing main supply
voltage. The rating plate is attached to the bottom
of the unit.
Place the charger on a level surface.
- To protect the device from over-heating, do not
obstruct the vent slots on the bottom of the unit.
Connect the power cable of the charger to the
power supply.
- The green LED on the left side of the device is
illuminated. This shows that the unit is ready for
operation; please refer to the section, "LED dis-
play of the charger".
Light Guide/Handpiece:
Never insert the handpiece in the charger unless the
battery is inserted in the handpiece first!
Steam Autoclave the light guide prior to first use.
Then insert the light guide in the handpiece until it
snaps into place.
Place the enclosed glare shield on the light guide.
Insert the battery:
Rotate the lid at the lower end of the handpiece
counterclockwise until it hits the stop, and then
remove the lid.
To avoid malfunction, make sure the battery is
properly inserted. Lay the handpiece upside down
and insert the rechargeable battery in a fluent
motion as directed by the arrow until you hear the
battery snap into place.
Replace the lid and rotate clockwise until it locks
into place.
4
In the event of malfunction, remove the rechargeable
battery from the device and re-insert as described
above.
Battery Charging:
Prior to the first use, place the handpiece in the
charger to fully load the new rechargeable battery.
- The yellow LED of the charger illuminates after
approximately 2 seconds; please see also section,
"LED Display of the Charger". The battery is fully
charged once the yellow LED is turned off.
- Note: Several cycles of charging and discharging
may be necessary before the optimum cure time
per charge is obtained.
LED Display of the Charger
Yellow LED
Green LED
LED
LED
Acoustical
green yellow
Signal
On
Off
---
On
Off
---
On
On
---
On
Flashing ---
On
Flashing Ongoing
beeps
Table Holder for the Handpiece
While performing a procedure, the handpiece can be
placed in a table holder.
no handpiece
charger
inserted in
ready for
the charger
operation
handpiece
charging
inserted in
has been
the charger
completed
handpiece
recharge-
inserted in
able
the charger
battery is
being
charged
handpiece
recharge-
inserted in
able
the charger
battery is
defective
handpiece
charging
inserted in
contacts
the charger
are wet