Descargar Imprimir esta página

3M ESPE Elipar FreeLight 2 Instrucciones De Manejo página 7

Ocultar thumbs Ver también para Elipar FreeLight 2:

Publicidad

Idiomas disponibles

Idiomas disponibles

If the light intensity is below 80% (fewer than 4 LEDs
are ON), check the light guide for contamination or
defects.
Proceed as follows:
- 1. Clean the light guide if contaminated; please
refer to the section on "Care", or
- 2. replace the light guide if defective, or
- 3. if 1. and 2. fail to improve the situation, please
contact 3M ESPE Customer Service or your
retailer.
Low Battery Charge Display
After frequent use the battery charge of the device may
drop to below approximately 10%, so that only a few
more exposures can be performed without re-charging.
Low battery charge is shown by several means:
- A short audible alarm is emitted 5x at the
completion of an exposure, and repeated
thereafter whenever any button is touched.
- The 5-second-LED of the handpiece begins to
flash.
The handpiece should be placed in the charger as
soon as possible to re-charge the battery.
Power-Down Mode
Once the handpiece is placed in the charger, all internal
functions and LEDs are automatically turned off as the
handpiece switches to power-down mode. This reduces
the power consumption of the rechargeable battery to
a minimal level. Outside the charger, the handpiece is
also switched to power-down mode if left unused for
approximately 10 minutes.
To terminate the power-down mode press one of
the two buttons.
- The power-down termination signal (two short
audible signals) is emitted indicating that the
6
Blue LEDs
80
60
40
%
handpiece is ready for operation: the handpiece
displays show the latest selected exposure mode
and time settings.
Acoustical Signals - Handpiece
An acoustical signal is emitted
• every time a button is pressed,
• every time the light is turned ON or OFF,
• after 5 seconds of exposure time have elapsed
(1 alarm signal), 2x after 10 seconds, 3x after
15 seconds.
Two acoustical signals are emitted
• every time the power-down mode is terminated by
pressing any of the buttons.
A 2-second-error signal is emitted, if
• the temperature control is activated,
• the rechargeable battery is discharged to an extent
that reliable polymerization is no longer ensured.
Acoustical Signals - Charger
Ongoing beeps are emitted when the handpiece is in
the charger and the charge contacts are wet.
Operating Errors
Error
The 5-second-LED
flashes and a short
signal is emitted
5x whenever a button
is touched or the light
is turned off.
The ongoing exposure The battery lacks sufficient
is interrupted (light
off signal is emitted)
followed by a 2 second-
error signal; the device
switches to power-
down mode and resists
further activation.
A 2 second-error
signal is emitted
upon pressing of the
Start button.
Cause
Solution
The residual battery charge
has dropped below 10%.
Place the handpiece in
the charger and re-charge
the battery.
charge.
Place the handpiece in
the charger and re-charge
the battery.
The temperature control
activated to protect the hand-
piece from overheating.
The handpiece can be used
again once it has cooled down.
Allow the handpiece to
cool down.
- The light may be success-
fully activated once the
handpiece has cooled
down.

Publicidad

loading