Trooper Scale
4.4.6 Lockout Switch Menu (Cont.)
Center
Gross
of Zero
Brutto
Procedure (Cont.)
With the Scale ON, press and hold the G/N/T/ MENU but-
ton until MENU is displayed. When you release the G/N/
T/ MENU button, CAL is displayed, then press PRINT/
UNITS button, until LOCSW is displayed.
Press G/N/T/ MENU button, LSTOFF is displayed. This
permits locking the Setup menu. OFF is unlocked, ON is
read only (locked). This menu is hidden if the CAL jumper
is off.
Press PRINT/ UNITS button, and select ON or OFF.
Press G/N/T/ MENU button, LrdOFF displayed. This per-
mits locking the Readout menu. OFF is unlocked, ON is
read only (locked).
Press PRINT/ UNITS button, and select ON or OFF.
Press G/N/T/ MENU button, LPtOFF is displayed. This
permits locking the Print menu. OFF is unlocked, ON is
read only (locked).
Press PRINT/ UNITS button, and select ON or OFF.
Press G/N/T/ MENU button, LCLOFF is displayed. This per-
mits locking the Calibration menu. OFF is un locked, ON
is hidden (locked). This menu is hidden if the LFT switch
is in the LFT position.
Press PRINT/ UNITS button, and select ON or OFF.
Press G/N/T/ MENU button to end this block, SAVE is
displayed.
Press G/N/T/ MENU , Quit is displayed.
Press PRINT/ UNITS button to go to CAL or press G/N/T/
MENU button, Scale returns to a weighing mode.
NOTE: At this point, the Lock Switch must be set in order
to lock out the menus.