Trooper Scale
5. CALIBRATION AND SEALING (Cont.)
Cal in kg:
TR3RS
TR6RS
TR15RS
TR30RS
Cal in lb:
TR3RS
TR6RS
TR15RS
TR30RS
Procedure (Cont.)
TURNING SOFTWARE LOCK OFF
Span cal choices
Linearity cal (fixed)
1, 2, 3kg
2, 4, 6kg
5, 10, 15kg
10, 20, 30kg
Span cal choices
Linearity cal (fixed)
2, 4, 6 lb
5, 10, 15 lb
10, 20, 30 lb
20, 40, 60 lb
For span calibration, there are a number of choices for
calibration. The highlighted number is the default.
To change the span calibration value read the setup sec-
tion of the menu on page 10. In the Setup menu, change
the CP selection to the desired value.
When the desired selection has been chosen and the
masses are available, you are ready to begin the calibra-
tion routine.
Procedure
SPAN CALIBRATION
2 & 3 kg
4 & 6 kg
10 & 15 kg
20 & 30 kg
4 & 6 lb
10 & 15 lb
20 & 30 lb
40 & 60 lb
With the Scale ON, press and hold the G/N/T/ MENU
button until MENU is displayed. When you release t h e
G/N/T/ MENU button, CAL is displayed. If the display
shows SETuP, read the section on turning off the Calibra-
tion Lock Switch or check the settings for legal for trade.
Press G/N/T/ MENU button, SPAN Y is displayed.
Press G/N/T/ MENU button, -C- is displayed. The scale
MUST be stable during this period and is establishing a
zero point. After a few seconds, the requested weight
value is displayed.
EN-23