PoolDose | pH · ORP · Chlorine
14. ALARMS
Alarm
Level
*only active measures
Out of Range measure
OFA First Alarm
(time >70%)
OFA Second Alarm
Flow Rate
Calibration Function
System Error
Alarm measure (*
1
)
WiFi Menu is not present in the
advanced menu, but WiFi card
is present
Error Measure
(*
Measure Alarm Ranges)
1
n
1
Temp. Measure min
2
Temp. Measure Max
3
pH Measure
min
4
pH Measure
Max
5
ORP Measure min
6
ORP Measure Max
7
CL Measure
min
8
CL Measure
Max
0000138994
Display
Lev Low
ALARM BAND
OFA_1
(time 100%)
OFA_2
Flow
CALIBRATION ERR
ParameterS error
High Measure
Low Measure
WiFi menu is not
present
Wrong measure
Item
Actions to do
- Push
- Restore Product tank
- Replace or check the measure probe
- Push
- Restore measure
- Press
press
- Press
press
- Restore Flow Rate
- Restore Probe or Buffer solution and repeat
calibration procedure
- Press
- Broken Unit
- Adjust the chemical concentration
- Switch off the unit, keep pressed
switch on.
- Please, reset the calibration parameter, into
Advanced Menu click to Reset CAL Item.
+ 10°C
+ 38°C
6 pH
8 pH
+ 600 mV
+ 800 mV
0,50 ppm
2 ppm
rev. 3.2
EN
to open Alarm Relay
to open Alarm Relay
for 3 seconds to reset or
for 5 seconds to reset
for 3 seconds to reset or
for 5 seconds to reset
to restore Default parameter
Limits
and
16