MESSAGE
Set clock.
No saved data.
Set clock.
Error Messages
ERROR #
MESSAGE
Please insert new
01
batteries.
Device memory error.
11
Cannot read data.
Device memory error.
12
Cannot erase data.
Device memory error.
13
Cannot write data.
SD memory error.
21
Cannot read data.
SD memory error.
22
Cannot erase data.
SD memory error.
23
Cannot write data.
MESSAGES
CAUSE
The battery power has run low
and the Date and Time setting on
the unit has disappeared.
The battery power has run low
and the ECG data stored on the
unit has disappeared or the clock
has been reset.
CAUSE
The batteries are worn out.
An error occurred while trying to
read ECG data stored in unit
memory.
An error occurred while trying to
erase ECG data stored in unit
memory.
An error occurred while trying to
save ECG data to unit memory.
An error occurred while trying to
read ECG data stored on the SD
memory card.
An error occurred while trying to
erase ECG data stored on the SD
memory card.
An error occurred while trying to
save ECG data on the SD
memory card.
52
CORRECTION
Insert new batteries and reset the
Date and Time from the SET menu.
Insert new batteries and reset the
Date and Time and other settings.
CORRECTION
Replace both batteries with new
ones.
Contact
Omron Healthcare
Customer Service
at 1-800-829-6427.
Format the SD memory card.
Refer to page 22.
* Copy any important data to a PC
before formatting the SD memory
card as all data on the card will be
erased.
Use a new/different SD memory
card. (Always turn off the unit
before removing or inserting SD
memory cards.)