39
Technical information for set-up
• The product is ready for operation as soon as it is removed from the packaging.
• Check the device visually for any external damage.
• Do not set up the device if there is any visible damage.
• Place the product on a level and solid surface, e.g. table.
• Insert the mains cable into the mains connector (B2) and connect to the socket
(voltage supply)
•
For the proper operation of the device and its connection to the electricity grid,
please use a country-specific mains plug adapter (not included in the delivery)
that meets the specifications of the device.
• Connect the product to a voltage supply in line with the specifications
• WARNING: To reduce the risk of an electric shock, this device may only be
connected to a power supply WITH A protective ground connector.
• Set the device up so that, in an emergency, the mains cable can be pulled out
by the patient or the operator during treatment.
• Do not place the device on cloths, blankets, duvets etc.
due to the risk of overheating
• Do not cover the ventilation slots (B1) on the device so as to avoid
overheating. Do not stack devices and do not use the device as a
storage surface
• Remove the dummy connector from the connection to be used (A7)
and connect the cuff
• Connect the cuffs to the device (A7) and put on
• The following standard setting will appear:
Speed (filling the cuffs): >> high (A1)
Interval (pause period between the compression cycles): 15 seconds (A2)
Pressure (Compression pressure): 20 mmHg (A3)
Time (Treatment period): 15 minutes (A4)
• After the desired compression pressure has been set (A3) the therapy
will begin after pressing the (A6) ("Start") button
lympha-mat
DIGITAL