Technical Specifications - MEGABRAS MD-5060e Manual De Usuario

Megóhmetro digital
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

10. Technical specifications

Test voltages
Maximum resistance
reading
Short circuit current
Test voltages accuracy
Equipment basic accuracy : 5% of reading ± 3 digits
Advanced features
Built-in printer
Serial data output
Built-in chronometer
Environmental protection
Safety class
Overvoltage category
E.M.C.
Electrostatic immunity
Electromagnetic irradiation
immunity
Power supply
Battery charger
Operating temperature
range
Storage temperature range : -25°C to +65°C
: 0.5 kV - 1 kV - 2.5 kV - 5 kV directly, one button
selectable. 0.5 kV to 5 kV in 500 V steps. DC,
negative.
: 5 TΩ @ 5 kV
: 1.5 ± 0.5 mA
: ± 3% of nominal test voltages @ R ≥ 10 GΩ
(1 MΩ to 500 GΩ at any test voltage)
: Automated Polarization Index computing
Automated Dielectric Absorption Ratio computing
: Prints elapsed time, actual voltage and resistance
measured each 15 seconds (offered as an optional
feature).
: RS232 @ 4800 bps. Suitable for data collection in
an external serial printer, computer or data-logger.
: Shows elapsed time in mm:ss format. Count starts
automatically for each measurement.
: IP54 (with closed lid).
: Meets the requirements of IEC 61010-1:1990,
IEC 61010-1:1992 amendment 2.
: EN 61010-1 to 600 V CAT. III
: In accordance with IEC 61326-1.
: In accordance with IEC 1000-4-2.
: In accordance with IEC 61000-4-3.
: Internal rechargeable 12 V - 2.3 Ah battery.
: 100 - 240 V~
: -5°C to +50°C
24