6
Tubing:
1. Disconnect the tubing from the unit.
2. Tubing should be rinsed thoroughly after every use by running hot tap water through it followed by a solution of one part vinegar to three
parts hot water.
3. Rinse with hot tap water and air dry.
4. Keep the outer surface of the tubing clean by wiping with a clean, damp cloth.
Carrying Case:
1. Wipe the case using a clean cloth dampened with detergent and/or disinfectant.
MAINTENANCE
Inspect suction tubing and collection bottle for leaks, cracks, etc. before each use.
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable internal components. If service is
required, return unit to a qualified Laerdal Medical provider or an authorized service center. Opening or tampering with the unit will
void warranty.
Filter Changing:
1. Change bacteria filter if overflow occurs or every two months, whichever comes first.
2. Remove filter by disconnecting it from suction unit and lid assembly.
3. Replace with a clean Laerdal bacteria filter (non-sterile) 88 49 01 (12/pack) and remount to suction unit and lid. Additional filters may be
purchased from your authorized Laerdal Medical provider.
NOTE– Do not substitute any other material for this bacteria filter. Substitution may lead to contamination or poor performance; use only Laerdal filters.
TROUBLESHOOTING
NOTE– Your suction unit contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO THE HOME MEDICAL
EQUIPMENT PROVIDER WHERE YOU PURCHASED IT OR TO Laerdal Medical, please take a few moments to check for these possible causes:
Problem
Unit does not turn on.
(Green external power indicator
should be illuminated when
power is applied.)
Pump runs,
but no vacuum.
Low vacuum.
Battery will not hold a charge.
(Charge indicator should be
illuminated if battery is con-
nected during charge mode.)
Battery seems not to hold a
charge.
100-405-00
E n g l i s h
DANGER
Action
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
3. If running from an internal battery, ensure that your unit has a
battery installed.
4. If battery is installed, check that it is fully charged.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks, leaks, or occlusions.
3. Ensure that flow shut-off is not activated.
4. Check for leaks or cracks in bottle assembly.
1. Use vacuum adjustment knob to increase vacuum level.
2. Check system for leaks.
3. Push vacuum adjustment knob and then release.
1. Ensure that unit is equipped with an internal battery by contacting
your Laerdal Medical equipment provider.
2. Verify that charge light turns on.
3. Check electrical connections during charging.
4. Ensure wall outlet is live by plugging in a lamp.
1. Perform the following test to determine if battery replacement is
necessary:
a. Charge battery as directed.
b. Disconnect charging accessory and operate the LCSU at free flow
(no suction load and tubing unobstructed) for 20 minutes.
If LCSU stops before completing the 20 minutes, contact Laerdal to
arrange for battery replacement.
M a i n t e n a n c e / T r o u b l e s h o o t i n g