S p e c i f i c a t i o n s / C l a s s i f i c a t i o n s
SPECIFICATIONS/CLASSIFICATIONS
Size
88 00 20, 88 00 30, & 88 00 40 Series . . . . . . . . . . . . . . . . .9.0 x 7.0 x 8.0 (22.9 x 17.8 x 20.3)
Weight
88 00 20 & 88 00 30 Series . . . . . . . . . . . . . . . . . . . . . . . . .6.3 (2.9)
88 00 40 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.8 (1.7)
Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240 V AC 50/60 Hz 0.75 A max; 12 V DC, 33 W max
Internal Rechargeable Battery
88 00 20 & 88 00 30 Series . . . . . . . . . . . . . . . . . . . . . . . . .Factory Equipped
88 00 40 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Not Factory Equipped (provider-installed option)
Vacuum Range
88 00 20, 88 00 30, & 88 00 40 Series . . . . . . . . . . . . . . . . .80 to 550 mm Hg
Air Flow @ pump inlet:
. . . . . . . . . . . . . . . . . . . . . . . . .27 LPM (free flow) typical (may be less when running from internal battery)
Collection Bottle Capacity
88 00 20 & 88 00 40 Series . . . . . . . . . . . . . . . . . . . . . . . . .800 ml (cc) Disposable
88 00 30 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1,200 ml (cc) Reusable
Environmental Conditions
Operating Temperature Range . . . . . . . . . . . . . . . . . . . . . . .32°F (0°C) - 104°F (40°C)
Operating Relative Humidity . . . . . . . . . . . . . . . . . . . . . . . .0-95%
Operating Atmospheric Pressure . . . . . . . . . . . . . . . . . . . . . .10.2 Psi (70 kPA) - 15.4 Psi (106 kPA)
Storage & Transport Temperature Range . . . . . . . . . . . . . . . . .-40°F (-40°C) - 158°F (70°C)
Storage & Transport Relative Humidity . . . . . . . . . . . . . . . . .0-95%
Storage & Transport Atmospheric Pressure . . . . . . . . . . . . . . .7.3 Psi (50 kPA) - 15.4 Psi (106 kPA)
Warranty
88 00 20 & 88 00 30 Series . . . . . . . . . . . . . . . . . . . . . . . . .Two-years limited, excluding internal battery and collection bottle
88 00 40 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Two-years limited, excluding collection bottle
Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90-day
Approvals
88 00 20, 88 00 30, & 88 00 40 Series . . . . . . . . . . . . . . . . .IEC 601-1; CAN/CSA-C22.2 No. 601.1-M90; UL 2601-1, CE EN 60601-1-2
88 00 20 & 88 00 30 Series meets RTCA/D-160D . . . . . . . . . .DO-160D - section 20 Category U
Equipment Classifications
With respect to protection from electric shock . . . . . . . . . . . .Class I and internally powered
Degree of protection against electric shock . . . . . . . . . . . . . .Type BF Applied Parts
Degree of protection against ingress of liquids . . . . . . . . . . . .IPX2 and ordinary power supply
Mode of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Intermittent Operation: 30 minutes on, 30 minutes off
Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or nitrous oxide.
ISO Classification
88 00 20 & 88 00 30 Series only - Electrically powered medical suction equipment for field and transport use according to ISO 10079-1 : 1991
High Flow/High Vacuum
88 00 40 Series - Electrically powered medical suction equipment for non-transport use according to ISO 10079-1 : 1991
H x W x D inches (cm)
lb. (kg)
DO-160D - section 21 Category L
For battery operation only
Airline use
E n g l i s h
7