Connection Ports Of The Titronic ® 300 Piston Burette; Back Panel Of The Titronic; Connecting The Piston Burette - Combination With Accessories And Additional Devices; Connecting A Printer - Xylem SI Analytics TITRONIC 300 Manual De Instrucciones

Burette a piston
Ocultar thumbs Ver también para SI Analytics TITRONIC 300:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

60
2.3 Connecting the piston burette – Combination with Accessories and Additional
Devices
2.3.1

Back panel of the TITRONIC

Fig. 3
2.3.2
Connection ports of the TITRONIC
The burette TITRONIC® 300 offers the following interfaces:
Two RS232 interfaces, 4-channel (Mini-DIN):
1) For connection of a weighing balance and other devices from von SI Analytics (burettes,etc.)
2) RS1 for connection to the PC
3) One USB-A ("Host") interface for connecting USB devices such as a keyboard, printer, manual control
unit, USB memory stick and USB hub
4) USB-B interface (mini USB) for connection to a PC
5) Connection of the external power supply TZ 1853
6) On/Off switch
2.3.3

Connecting a printer

Printers with a USB interface are to be connected to one of the two USB-A interfaces. These printers have to
feature HP PCL emulation (3, 3 enhanced, 5, 5e, no GUI and GDI printers).
2.3.4

Connecting a USB device (manual controller, keyboard, memory device, hub)

The following USB devices can be connected to the USB-A interfaces:
PC-keyboard
TZ 3880 manual controller
Printer
USB storage devices, e.g. USB sticks
USB hub
USB barcode scanners
For connecting more than one USB device, a USB hub (e.g. TZ 3830) is required.
2.3.5

Connection of analytical balances

Analytical balances are to be connected to the RS232-2 port using an appropriate cable.
®
300 piston burette
1
2
3
4
5
6
®
300 piston burette

Publicidad

Tabla de contenido
loading

Tabla de contenido