MAINTENANCE
Regular maintenance is essential for good operations over a long period. Follow the advice given here
below.
Before any operation, make sure the system is unplugged!
Replacing Fuses
This operation must be performed only by qualified personnel. If the intervention of a technician
is not possible, send the instrument to the manufacturer for proper maintenance.
Remove the unit from the mounting bracket
-
-
Unscrew the case screws and remove the back panel
Replace the blown fuse with a new one of the same type (size and value)
-
If the fuse blows again, send the instrument to the manufacturer for repair
-
-
Close the back panel
Maintenance of the Redox Electrode
Typically, it is recommended to clean the electrode when the response is slow or measurements are not
reliable, and when it has been used for a long time, especially in aggressive solutions, pollutants, very
acidic or very alkaline environments.
STEIEL provides a kit of solutions for the cleaning and storage of pH and redox electrodes.
The kit includes three solutions:
Solution A: dip the electrode in this solution for cleaning it
-
Solution B: use this solution to rinse the electrode, before and after cleaning
-
-
Solution C: storage solution to be used for filling the electrode protective cap when the electrode is not
used (wintering, closure of the plant)
EF117_r1.0 – 05/2018
Input protection, F1
Output protection, fuse inside the in-line fuse-holder
p. 14