Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations
on power
supply input
lines
IEC 61000-4-11
Tesmed copyright - it is absolutely forbidden the partial or total reproduction of any present content
IEC 60601
Test level
±8kV direct &
indirect contact;
±15kV air
discharge
±2 kV for power
supply lines
± 1 kV line(s) to
line(s)
<5% UT
(>95% dip in
UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in
UT)
for 5 sec
85
Compliance
Electromagnetic
level
environment-
guidance
±8kV direct
Floors should be
& indirect
wood, concrete or
ceramic tile. If floors
contact;
±15kV air
are covered with
discharge
synthetic material,
the relative humidity
should be at least
30%
Not
not applicable
applicable
(for INTERNALLY
POWERED ME
EQUIPMENT)
Not
not applicable
applicable
(for INTERNALLY
POWERED ME
EQUIPMENT)
Not
not applicable
applicable
(For INTERNALLY
POWERED ME
EQUIPMENT