Compatibility
The Skin IQ MCM is designed to fit on a pressure redistribution surface that is 203.2 cm - 213.4 cm
(80 - 84 in) long by 88.9 cm - 91.4 cm (35 - 36 in) wide by 17.8 cm (7 in) high.
Consult product labeling for the pressure redistribution surface and / or bed frame for compatibility.
Risks and Precautions
Transfer
This product is not intended for use as a transfer device.
Duration of Use
Recommended duration of use for single patient is not more than 60 days for patients
< 172.37 kg (380 lb).
Duration of use for patients weighing 172.37 kg (380 lb) - 227 kg (500 lb) is not to exceed more
than 30 days.
However, patient specific duration of use may vary. Clinical conditions such as, but not limited to,
incontinence, skin condition, nutrition status, medications, mobility, weight, or etiology need to be
considered when assessing duration of use for the Skin IQ MCM.
Height
The Skin IQ MCM will increase the height of the pressure redistribution surface it is applied to by
approximately 6.35 mm (0.25 in).
Use With Other Devices
All Skin IQ MCM components are designed to be used as a single system device.
The included power supply should only be used with the Skin IQ MCM. The coverlet should only
be powered by the Skin IQ Power Supply Arjo part number 4103832.
Any attempt to connect and use the power supply with any other device, or
use any other brand or model of power supply other than Arjo part number
4103832 will result in improper operation of equipment, possibly leading to
increased risk of patient injury.
Patient Migration
Specialty surfaces have different shear and support characteristics than conventional surfaces and
may increase the risk of patient movement, sinking and / or migration into hazardous positions
of entrapment and / or inadvertent bed exit. Monitor patients frequently to guard against patient
entrapment.
EN
5