3.10.2.2 Equipment
Cotton wool
CAUTION If you use a lens cleaning cloth it must be dry. Do not use a lens
cleaning cloth with the liquids that are given in section 3.10.2.1 above. These
liquids can cause material on the lens cleaning cloth to become loose. This mate-
rial can have an unwanted effect on the surface of the lens.
3.10.2.3 Procedure
Follow this procedure:
1.
Soak the cotton wool in the liquid.
2.
Twist the cotton wool to remove excess liquid.
3.
Clean the lens one time only and discard the cotton wool.
WARNING Make sure that you read all applicable MSDS (Material Safety Data
Sheets) and warning labels on containers before you use a liquid: the liquids can
be dangerous.
CAUTION
Be careful when you clean the infrared lens. The lens has a delicate anti-re-
flective coating.
Do not clean the infrared lens too vigorously. This can damage the anti-reflec-
tive coating.
3.10.3 Decontamination and disinfection
The camera must be thoroughly cleaned, decontaminated and disinfected be-
fore shipping to our service department. No hazardous residues are allowed
on cameras. Such residues include—but are not limited to—chemical fire-ex-
tinguishing compounds, radioactivity, biohazardous materials, and residues
from chemical fires.
FLIR Systems reserves the right to charge the full cost for the decontamina-
tion and disinfection of contaminated cameras that are shipped to our service
department.