Operation
Self test
Warm-up time
60
During the self-test, all cells must be removed and the cell shaft cover closed.
Self test
System test
System test
Filter test
Lamp test
Wavelength calibration
Note
The result of the self-test can be viewed and printed with the [Info] function
key.
After switching on the photometer requires a warm-up time of 15 minutes.
Reproducibility of measurement data is restricted during the warm-up time.
Therefore, do not measure during the warm-up time.
During the warm-up time, a progress bar appears on the display next to the
date. The progress bar disappears as soon as the warm-up time is over.
The self-test includes:
04/16/07 9:52
– the test of the memory,
processor,
internal interfaces,
filter and lamp
– a calibration for each
wavelength
After the self-test is completed,
the main menu is displayed.
Progress bar during
warm-up time
®