What to do if ...
Self test failed.
Connected printer
does not print
84
Cause
– System test: Instrument defective
– Filter test: Instrument defective
– Wavelength calibration:
– Foreign particle in the cell shaft
– Lens smudged
– Instrument defective
Cause
– Printer not suitable
photoLab
Remedy
– Return instrument to service
department
– Return instrument to service
department
– Remove foreign object
– Clean the lens (see section 5.2.3
or section 6.1). If this happens
repeatedly, check the operating
conditions (see section 3.1)
– Return instrument to service
department
Remedy
– Connect a printer that can
interpret the printer control
language PCL-3