•
When servicing parts inside the Analyzer, ensure
that power to the Analyzer is OFF, and keep
hands clear of the reagent syringe assemblies.
The syringes are controlled by moving parts that
can pinch skin.
HARZARDOUS WASTE DISPOSAL
Hazardous reagents (ammonium persulfate
and phosphoric acid) are used in the Analyzer.
The waste stream from the instrument is acidic
and must be disposed of properly. Consult
your federal, state, and local government
regulations.
•
The UV lamp and the display screen contain
mercury and may be considered hazardous
material in your local area. Dispose of these
items in accordance with federal, state, or local
government regulations.
•
In the case of any broken or damaged UV lamp,
handle the remains in accordance with your
organization's toxic waste handling procedure and
dispose of them in accordance with federal, state,
or local government regulations.
This symbol indicates the European Union
Directive 2002/96/EC for waste electrical
and electronic equipment (WEEE) requires
disposing the Analyzer separately from
standard waste.
OPERATIONAL CAUTIONS
•
Ensure that the DI water Reservoir is full,
particularly when running samples with high TOC
or high salt concentrations. After running high
TOC or salt samples, always rinse the Analyzer by
running low-TOC DI water through a sample cycle.
•
To avoid false TOC readings and possible damage
to the Analyzer, always ensure that the sample
inlet is open and the DI water reservoir is filled
before starting analysis.
•
Operation of the Analyzer without the in-line filter
on the sample inlet line will damage the Analyzer
and void the warranty. To avoid damaging the
Analyzer, install the filter and replace the filter
element, as needed.
•
To avoid damage to the DI water pump, ensure
that the DI water reservoir is filled BEFORE
turning On the Analyzer.
•
To avoid false TOC readings and possible damage
to the Analyzer, always ensure that the sample is
flowing through the iOS System and the DI water
reservoir is filled before starting analysis.
•
For performance within specifications on ozonated
water systems, an Ozone Destruct Kit must be
purchased from Veolia and installed according to
instructions.
•
When cleaning the Analyzer, always apply liquid to
the cloth directly.
This symbol on the instrument indicates that
the product does contain restricted substances
included in China RoHS II. Please refer to the
following table.
Note: This product contains a UV Lamp (with Hg) which meets
the exemption defined in Annex III 4(a) of RoHS Directive
2011/65/EC.
M-Series Quick Start Guide DQS 77000-03 MUL Rev A
warnings
33