specification summary: m
Operating Specifications
Range
Precision
Accuracy
Sample Type
Display Readout
Calibration
Analysis Time
Sample Temperature
Ambient Temperature
Sample Pressure
On-Line Flow Rate
Instrument Sample Flow Rate
Analyzer Specifications
On-Line Inlet(s)
Outputs
Display
Power
Fuses
Dimensions
Weight
Enclosure Rating
Safety Certifications
Environment
Maximum Relative Humidity
Maximum Altitude
Pollution Degree
The UV lamp inside this product contains mercury and must be recycled or disposed of in accordance with local, state, and federal laws.
Hg
© 2023 Veolia. All rights reserved.
© 2023 Veolia. All rights reserved.
4
5310C
M5310C
Laboratory Analyzer
± 2% or ± 0.5 ppb, whichever is greater
Autosampler or
discrete grab sample,
and TOC removal grab
N/A
N/A
N/A
USB device port (1),
USB host ports (3),
Ethernet
7" WVGA 800x480 pixel, Color LCD w/ touch-screen
Replace with same type and size fuse:
T 1.6 A 250 VAC Fuse (Slow Blow), size 5 x 20 mm appliance inlet
H: 42.2 cm (16.6 in.);
W: 24.6 cm (9.7 in);
D: 40.0 cm (15.8 in)
10.0 kg (22.0 lb)
N/A
M5310C
On-Line Analyzer
4 ppb to 50 ppm
<1% RSD
On-line continuous
or discrete grab sample,
timed on-line,
TOC removal on-line,
TOC removal grab
3 significant digits
Typically stable for 12 months
2 minutes
5-60 °C (41-140 °F)
5-40 °C (41-104 °F)
>50 mL/min (for on-line mode)
0.5 mL/min
One stream,
or 2-stream inlet (option)
4-20 mA outputs (3); alarm outputs (4);
binary input (1); USB device port (1),
USB host ports (2); Ethernet TCP/IP
°
100 – 240 V~, 50 – 60 Hz, 100 VA
H: 54.9 cm (21.6 in);
W: 45.0 cm (17.7 in);
D: 26.5 cm (10.4 in)
16.2 kg (35.6 lb)
IP-45
ETL. CE, UKCA
0 - 95%, non-condensing
3,000 m (9,800 ft)
2
M5310C
Portable Analyzer
On-line continuous,
Autosampler, or discrete
grab sample, timed on-line,
TOC removal grab
100 psig
One stream
H: 39.5 cm (15.4);
W: 22.9 cm (9.0 in);
D: 46.4 cm (18.3 in)
9.8 kg (21.6 lb)
IP-21