21. EMC Information
The instrument fulfils the requirements according to EN 60601-1-2 regarding emission and immunity.
If sensitive electronic equipment is affected by the instrument, try to increase the distance to such equipment.
The charger should not be connected during measurements.
Guidance and manufacturer's declaration – Electromagnetic Emissions
Osseo 100 is intended for use in the electromagnetic environment specified below.
Emissions tests
RF emissions C1SPR11
RF emissions C1SPR11
Harmonic emissions IEC61000-3-2
Voltage fluctuations/flicker emissions IEC61000-3-3
Guidance and manufacturer's declaration – Electromagnetic Immunity Test Levels
Osseo 100 is intended for use in the electromagnetic environment specified below.
Immunity test
Electrostatic discharge (ESD)
Radiated RF EM fields
Proximity fields form RF wireless communications
equipment
Rated power frequency magnetic fields
Electrical fast transient/burst
Surges Line-to-line, Surges Line-to-ground
Conducted disturbances induced by RF fields
Voltage dips, Voltage interruptions and Electrical
transient condition along supply lines
Compliance
Electromagnetic environment – guidance
Group 1
Osseo 100 uses RF energy only for its internal
function.
Class B
Osseo 100 Rechargeable battery operated device.
Not applicable
Not applicable
EMC standard or
Professional healthcare facility environment
test method
IEC61000-4-2
± 8kV contact
± 2 kV ± 4 kV ± 8 kV ± 15 kV air
IEC61000-4-3
3 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
IEC61000-4-3
30 cm minimum separation distance from radio
transmitter
IEC61000-4-8
30 A/m
50 Hz or 60 Hz
IEC 61000-4-4
± 2kV
100 kHz repetition frequency
IEC 61000-4-5
± 0.5, ± 1 kV, ± 2 kV
IEC61000-4-6
3V
0,15 MHz – 80 MHz
6 V in ISM bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
IEC 61000-4-11
5% UT, 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT; 1 cycle
And 70 % UT; 25/30 cycles (50/60Hz)
Single phase: at 0°
0 % UT; 250/300 cycle (50/60 Hz)