11. ERROR MESSAGES AND REMEDIES
Error message
Error Code 1
Error Code 2
Error Code 3–4
Error Code 5–6
Error Code 7–8
Error Code 9–12
Error Code 101
Error Code 104
Error Code 105
If a warning message or error message is displayed, it can be closed via <OK>.
2119-2208-764120_IFU Evo_USA_PEMF_764120-GC-3015_Inhalt_v08.indd 68
2119-2208-764120_IFU Evo_USA_PEMF_764120-GC-3015_Inhalt_v08.indd 68
Cause
An active applicator was
removed during treatment.
The device temperature
is too high.
An application module has
reported an error.
Power supply malfunction
The controller has
reported an error.
An application module has
reported an error.
No applicator is connected.
An unknown applicator
was detected.
The charge level of the
B.Box Evo Battery Stand
is too low.
Troubleshooting
Reconnect the applicator and restart the
treatment.
Check the ambient conditions and allow
the device to cool down. If in doubt, please
contact customer service.
Please contact customer service.
Please check possible sources of error,
such as the B.Box Evo Power Supply or
B.Box Evo Battery Stand (see technical
data). If in doubt, please contact customer
service.
Please contact customer service.
Please contact customer service.
Please connect an applicator.
The applicator module is invalid, please
contact customer service.
The B.Box Evo Battery Stand is requires
charging, please connect the B.Box Evo
Power Supply to charge it.
68
19.06.2023 09:50:39
19.06.2023 09:50:39